2019-01-09

Compliance to DICOM Part 14.  Ability to clearly view medical images such as X-rays

As a leader in providing medical grade panel PCs to the healthcare industry, Wincomm is continuing to push the limits in creating the most effective and user-friendly solutions available. With this in mind, we are proud to announce that the WMP-24 inch series for medical imaging has added a new customizable function- DICOM 

proimages/news/Product_news/2015/20150430-1.jpgproimages/news/Product_news/2015/20150430-2.jpg

DICOM is the real luminance response of human eyes on the PC screen rendered in grayscale, so that an X-ray image can be developed and viewed more clearly. Wincomm’s DICOM design concepts include a Function Key to easily start up DICOM, automatic brightness fixing and accuracy, and having three color temperature modes for consistent preferences in different regions.

Picture: General screen VS DICOM

proimages/news/Product_news/2015/20150430-3.jpg

 Picture: Three kinds color temperatures modes

proimages/news/Product_news/2015/20150430-4.jpg

 

 

Picture: Function Key Design, press the button to operate DICOM, brightness will be fixed

proimages/news/Product_news/2015/20150430-5.jpg

 

Of the four medical display grade classifications, Wincomm’s DICOM Solution is the Clinical Display. This was chosen to provide affordability, color-accuracy, high-quality, and stable alternative solutions for clinics to enhance the quality of medical care via medical imaging (X-ray..).

proimages/news/Product_news/2015/20150430-6.jpg

 

Wincomm DICOM product shipments have been pre-calibrated and come with pre-set DICOM compliance reports. In addition, optional adjustment tools are available for purchase (including color sensor, dongle hard key, and system installer). Wincomm has the ability to customize DICOM since the WMP-24 inch LCD display specs are advanced enough to easily meet the stringent standards of DICOM.

proimages/news/Product_news/2015/20150430-7.jpg

 

Wincomm DICOM highlights

1. Compliance to DICOM Part 14 : Grayscale Standard Display Function, tolerance is less than 5%

2. Color temperature supporting: 5600K/6500K/9300K

3. Pre-set DICOM Compliance report of each finished product is enclosed

4. Calibration interval: 6 months

5. DICOM calibration tool (optional): Includes color sensor, dongle hard key, and system installer (execution programs, color sensor SDK, user manual)

 

 Everything, from the motherboard, BIOS, driver to the mechanical design was designed by Wincomm, so special customer requirements can be catered to and a steady supply provided over the long-term. The WMP-24’’ Series with customized DICOM function now available for sale. If you have any product or customization requirements, please contact Wincomm sales (mail: sales_support@wincomm.com.tw). For more product information, please visit the Wincomm website at https://www.wincommusa.com/

e-News Subscribe